U
SER
M
ANUAL
M
ONOCHROME
–
R
EV
J
–
05/2017
P
A G E
|
21
Manufacturer Info
(
ManufacturerInfo
) : Camera Part Number set by in factory in a 64Bytes String.
Read function (ASCII): “
r idnb
”;
Returned by the camera : <Part Number> String of 64 bytes (including “/0”)
Can’t be written
Serial Number
(
DeviceID
) : Camera serial Number set by in factory in a 64Bytes String.
Read function (ASCII): “
r deid
”;
Returned by the camera : Serial Number of the camera in a String of 64 bytes (including “/0”)
Can’t be written
Device User ID
(
UserDefinedName
) : Camera identifier set by the User in a 64Bytes String.
Read function (ASCII): “
r cust
”;
Returned by the camera : String of 64 bytes (including “/0”)
Write function (ASCII): “
w cust <idstr>
”
7.2
Device Privilege, Status and Reboot
GenCP address
ASCII
command
GenICam
command
Size R/W
Description
0x17040
lock
PrivilegeLevel
4
RW
Read:
-
0 : Factory
-
1 : Advance User
-
2 : User
Write :
-
1 : change mode from factory to
AdvanceUser
-
2 : change mode to User
-
Other: key to unlock the camera
0x17048
stat
Status
4
RO
Camera Status; bit set when :
Bit0
:no trigger during more than 1s
Bit1
: trigger too fast
Bit2
:
reserved
Bit8
: overflow occurs during FFC calibration
Bit9
: underflow occurs during FFC calibration
Bit16
: hardware error detected
0x17050
boid
BoardID
32
R
Unique Board Identification. Written by the
camera manufacturer or test bench
0x17070
bost
BoardStatus
16
R
Give the status of the board. Written by the
camera manufacturer or the test bench
0x17080
boot
RebootCamera
4
WO
Reboot the camera with a command
-
1 restart the camera (like a power cycle)
-
2 restart only camera application (bypass
upgrade application)