Problem:
Program Switching is disabled.
Reason: Solution switching is disabled when you edit solution files, and when you
export Solutions. Switching is re-enabled when you save or load a Solution.
Solution:
Save the current solution, or go to the Select Solution panel and load a
solution.
Note:
Loading a Solution in the Run panel does not re-enable switching.
Problem:
The Solution file will not load.
Reason 1: The Solution does not match the current BOA Spot (resolution or model).
Solution 1:
Try loading the Solution in the Emulator and see what model is displayed. If
you have trouble, try changing the resolution or model in the Emulator before loading.
Reason 2: If the Solution file will not load in the Emulator, it may be corrupted.
Solution 2:
Create a new Solution, or use an earlier version of this solution (if
available) from before the loading problem began.
Problems Running BOA Spot Solutions
Problem:
Always outputs a Fail pulse, even when the Monitor shows a Pass result.
Reason: The pulse offset may need to be increased, to allow for the inspection time.
Solution:
Please refer to the formula on page
, under “Pulse Setting”.
55
Problem:
Display controls are disabled in the Run panel. Cannot change the display.
Reason: These controls are disabled when image logging is enabled.
Solution:
Disable image logging to allow changing the display settings.
Problem:
I cannot reconnect to the BOA Spot. I keep getting a message to run
iAssistant and clean the system. But running iAssistant and cleaning is not fixing this
problem.
Reason: Either the BOA Spot or the PC may be “stuck” with a running process that
does not quit when is should. Sometimes using sensors with different firmware versions
(BOA, BOA2XA, BOA Spot) can cause this problem.
Solution:
1) Disable auto load (if possible) and reboot the BOA Spot. 2) Reboot the
PC. 3) Run iAssistant again, and click the “Clean” button. 4) Reboot the PC again.
5)
_
Now reconnect to the BOA Spot. New support files should be installed.
Problem:
History Log disabled in the Run panel.
Reason: The History Log is disabled by default to save on memory.
Solution:
History Log is enabled from the Connections panel.