TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
9.1.3.1 Signal I/O Diagnostic
Table 9-4: Diagnostic Mode - Signal I/O
No
Signal
Control
Description
0
DSP_BROWNOUT
NO
Display brownout is used to keep the display from
getting corrupted during low line voltage conditions.
Circuitry on the Status/Temp board senses low line
voltage and sets this bit. The CPU reads this and
generates the BROWNOUT_RST signal described
below.
1
EXT_ZERO_CAL
NO
Shows state of status input bit to cause the 6200A to
enter Zero Calibration mode. Use to check external
contact closure circuitry.
2
EXT_SPAN_CAL
NO
Shows state of status input bit to cause the 6200A to
enter the Span Calibration mode. Use to check
external contact closure circuitry.
3
SPAN_VALVE
YES
Switches the IZS Zero/Span valve. Use this bit to test
the valve function.
4
CAL_VALVE
YES
Switches the IZS Sample/Cal valve. Use this bit to
test the valve function.
5
DARK_SHUTTER
YES
Energizes the shutter. Use this bit to test the shutter
function which will block mechanically the UV lamp
light source.
6
RCELL_ HEATER
NO
Shows the status of the reaction cell heater. This has
the same function as the LED in the power supply
module.
7
IZS_HEATER
NO
Shows the status of the IZS permeation tube heater.
This has the same function as the LED in the power
supply module.
8
ELEC_TEST
YES
Turns on electric test bit in preamp. Should be used
for troubleshooting preamp circuit. We recommend
you use the ELEC TEST button in the DIAG menu to
operate electric test.
9
OPTIC_TEST
YES
Turns on optic test bit in preamp. Should be used for
isolating PMT detector system from the rest of the
preamp circuit. We recommend you use the OPTIC
TEST button in the DIAG menu to operate optic test.
(table continued)
9-13
Содержание 6200A
Страница 1: ...MODEL 6200A TRS ANALYZER MANUAL ADDENDUM TELEDYNE ANALYTICAL INSTRUMENTS 03655 Rev A August 2000 ...
Страница 9: ...TAI Thermal Converter Operator Manual Page 5 FIGURE 2 1 BASIC CONFIGURATION ...
Страница 10: ...TAI Thermal Converter Operator Manual Page 6 FIGURE 2 2 WITH IZS OPTION ...
Страница 11: ...TAI Thermal Converter Operator Manual Page 7 FIGURE 2 3 WITH M702 CALIBRATOR OPTION ...
Страница 13: ...TAI Thermal Converter Operator Manual Page 9 FIGURE 2 4 TRS CONVERTER LAYOUT ...
Страница 15: ...TAI Thermal Converter Operator Manual Page 11 FIGURE 2 5 PNEUMATIC CONNECTIONS ...
Страница 22: ...TAI Thermal Converter Operator Manual Page 18 FIGURE 3 1 TRS WIRING DIAGRAM ...
Страница 25: ...TAI Thermal Converter Operator Manual Page 21 FIGURE 4 1 DILUTER FLOW BLOCK ASSEMBLY ...
Страница 27: ......
Страница 35: ......
Страница 40: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 1 Removal of Shipping Screws 2 3 ...
Страница 41: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 2 Rear Panel 2 4 ...
Страница 42: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 3 Rear Panel Pneumatic Recommendations 2 5 ...
Страница 47: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 4 Front Panel 2 10 ...
Страница 48: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 5 Assembly Layout 2 11 ...
Страница 51: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 2 14 ...
Страница 58: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 4 1 6200A Sulfur Dioxide Analyzer 4 3 ...
Страница 65: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 5 2 Setup Menu Tree 5 2 ...
Страница 89: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 6 1 Permeation Tube Installation 6 8 ...
Страница 107: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 6 26 ...
Страница 110: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 7 1 Model 6200A Calibration Setup 7 3 ...
Страница 136: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 1 Replacing the Particulate Filter 8 3 ...
Страница 140: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 2 Reaction Cell 8 7 ...
Страница 141: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 3 Pneumatic Diagram 8 8 ...
Страница 175: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 1 CPU Board Jumper Settings 9 32 ...
Страница 184: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 3 V F Board Dip Switch Settings 9 41 ...
Страница 189: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 4 Power Supply Module Layout 9 46 ...
Страница 190: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 5 Electrical Block Diagram 9 47 ...
Страница 193: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 6 Flow Pressure Sensor 9 50 ...
Страница 194: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 7 SO2 Sensor Module 9 51 ...
Страница 195: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 8 SO2 Sensor Module 9 52 ...
Страница 198: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 9 PMT Cooler Subsystem 9 55 ...
Страница 204: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 11 Kicker Leak Check 9 61 ...