Table 7-7: IZS or Z/S Valves Modes with Remote Contact Closure
Ext Zero CC
Ext Span CC
Operation
Contact Open
Contact Open
State when in SAMPLE mode, normal monitoring.
Contact Open
Contact Closed
Span check or calibrate
*
Contact Closed
Contact Open
Zero check or calibrate
*
* Calibrate only if Dynamic Calibration is enabled (see Table 9-5).
7.8 EPA Protocol Calibration
If the 6200A is to be used for EPA SLAMS monitoring, it must be calibrated in accordance with
the instructions in this section.
In order to insure that high quality, accurate measurements are obtained at all times, the 6200A
must be calibrated prior to use. A quality assurance program centered on this aspect and
including attention to the built-in warning features of the 6200A, periodic inspection, regular
zero/span checks and routine maintenance is paramount to achieving this.
In order to have a better understanding of the factors involved in assuring continuous and
reliable information from the 6200A, it is strongly recommended that Publication No. PB 273-
518 Quality Assurance Handbook for Air Pollution Measurement Systems (abbreviated, Q.A.
Handbook) be purchased from the NTIS (phone 703-605-6000). Special attention should be paid
to Section 2.9 which deals with fluorescence based SO
2
analyzers and upon which most of this
section is based. Specific regulations regarding the use and operation of ambient sulfur dioxide
analyzers can be found in 40 CFR 50 and 40 CFR 58. Both publications are available from the
U.S. Government Printing Office (phone 202-512-0327).
7.8.1 Calibration of Equipment
In general, calibration is the process of adjusting the gain and offset of the 6200A against some
recognized standard. The reliability and usefulness of all data derived from any analyzer depends
primarily upon its state of calibration. In this section the term dynamic calibration is used to
express a multipoint check against known standards and involves introducing gas samples of
known concentration into the instrument in order to adjust the instrument to a predetermined
sensitivity and to produce a calibration relationship. This relationship is derived from the
instrumental response to successive samples of different known concentrations. As a minimum,
three reference points and a zero point are recommended to define this relationship. The true
values of the calibration gas must be traceable to NIST-SRM's.
7-9
Содержание 6200A
Страница 1: ...MODEL 6200A TRS ANALYZER MANUAL ADDENDUM TELEDYNE ANALYTICAL INSTRUMENTS 03655 Rev A August 2000 ...
Страница 9: ...TAI Thermal Converter Operator Manual Page 5 FIGURE 2 1 BASIC CONFIGURATION ...
Страница 10: ...TAI Thermal Converter Operator Manual Page 6 FIGURE 2 2 WITH IZS OPTION ...
Страница 11: ...TAI Thermal Converter Operator Manual Page 7 FIGURE 2 3 WITH M702 CALIBRATOR OPTION ...
Страница 13: ...TAI Thermal Converter Operator Manual Page 9 FIGURE 2 4 TRS CONVERTER LAYOUT ...
Страница 15: ...TAI Thermal Converter Operator Manual Page 11 FIGURE 2 5 PNEUMATIC CONNECTIONS ...
Страница 22: ...TAI Thermal Converter Operator Manual Page 18 FIGURE 3 1 TRS WIRING DIAGRAM ...
Страница 25: ...TAI Thermal Converter Operator Manual Page 21 FIGURE 4 1 DILUTER FLOW BLOCK ASSEMBLY ...
Страница 27: ......
Страница 35: ......
Страница 40: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 1 Removal of Shipping Screws 2 3 ...
Страница 41: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 2 Rear Panel 2 4 ...
Страница 42: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 3 Rear Panel Pneumatic Recommendations 2 5 ...
Страница 47: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 4 Front Panel 2 10 ...
Страница 48: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 2 5 Assembly Layout 2 11 ...
Страница 51: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 2 14 ...
Страница 58: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 4 1 6200A Sulfur Dioxide Analyzer 4 3 ...
Страница 65: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 5 2 Setup Menu Tree 5 2 ...
Страница 89: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 6 1 Permeation Tube Installation 6 8 ...
Страница 107: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G INTENTIONALLY BLANK 6 26 ...
Страница 110: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 7 1 Model 6200A Calibration Setup 7 3 ...
Страница 136: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 1 Replacing the Particulate Filter 8 3 ...
Страница 140: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 2 Reaction Cell 8 7 ...
Страница 141: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 8 3 Pneumatic Diagram 8 8 ...
Страница 175: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 1 CPU Board Jumper Settings 9 32 ...
Страница 184: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 3 V F Board Dip Switch Settings 9 41 ...
Страница 189: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 4 Power Supply Module Layout 9 46 ...
Страница 190: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 5 Electrical Block Diagram 9 47 ...
Страница 193: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 6 Flow Pressure Sensor 9 50 ...
Страница 194: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 7 SO2 Sensor Module 9 51 ...
Страница 195: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 8 SO2 Sensor Module 9 52 ...
Страница 198: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 9 PMT Cooler Subsystem 9 55 ...
Страница 204: ...TAI Model 6200A SO2 Analyzer Operator Manual 02164 Rev G Figure 9 11 Kicker Leak Check 9 61 ...