Teledyne API Model 465L O3 Monitor Operation Manual
48
Teledyne API Model 465L O
3
Monitor Manual
05509G DCN6806
7.2.4.
VIEW Menu
The VIEW menu allows the operator to view various measurement parameters in
real-time. This can be useful for diagnosing various instrument or system problems.
Note that these values are updated in real-time as they are repeatedly scrolled on the
display.
Table 7-5: VIEW Menu Parameters
Parameter
Meaning
Units
Normal Range*
MEAS
UV detector reading, measure cycle
mV
250 – 1230
REF
UV detector reading, reference cycle mV
250 – 1230
PRESSURE
Sample pressure
psia
9.0 – 14.9
LAMP
TEMP
UV lamp temperature
ºC
50.0 – 52.0
SAMPLE
TEMP
Sample temperature
ºC
5.0 –50.0
FLOW
Sample Gas Flow Rate
cc/min
640 – 960
*After initial 20 minute warm-up period.
7.2.5.
CONFIG Menu
The CONFIG menu displays specific information related to the instrument’s
configuration. No changes are made within this menu.
Table 7-6: CONFIG Menu Parameters
CONFIG
Meaning
S/N
Serial number of 465L. (Should match
the serial number tag on the rear panel
of the instrument).
COMM ADDRESS
Displays the MODBUS slave ID for
the instrument.
STREAM COUNT
Displays the number of streams
configured. (Always 1 for instruments
without the Multi-stream Option).
REV
Revision of instrument firmware
BREV
Revision of sensor firmware
BOOT
Revision of bootloader firmware