7.2.3.
DIAG Menu
The DIAG menu (see Figure 7-2) contains functions that are useful for testing and
configuring external equipment that may be connected to the rear panel I/O of the
instrument.
7.2.3.1.
Analog Step-Test
When placed in this mode, the instrument will automatically step the analog output
(See Section 4.4.2 for details on the analog output connection) through 5 points
from 0 to 100 % and display the current value on the front panel. The operator can
suspend the automatic cycling and manually control the stepping by pressing either
one of the
UP
/
DOWN
arrow buttons. To resume automatic cycling, exit the
function by pressing
CFG
and re-enter by pressing
ENT
.
The 5 points will correspond to the following analog output values depending on
whether the output is configured for 0-5V or 4-20mA:
Table 7-3: Analog Step Test Values
Point
0-5V Output
4-20 mA Output
0%
0.00 V
4 mA
25%
1.25 V
8 mA
50%
2.50 V
12 mA
75%
3.75 V
16 mA
100%
5.00 V
20 mA
Press
CFG
to exit from the Analog Step-Test function.