Teledyne API Model 465L O3 Monitor Operation Manual
05509G DCN6806
Teledyne API Model 465L O
3
Monitor Manual
45
Signal Name
Rear Panel Label
Function
XRELAY 12
STRM 4 - 3
STREAM 4 HI-HI ALARM
XRELAY 13
STRM 5 - 1
STREAM 5 ACTIVE
XRELAY 14
STRM 5 – 2
STREAM 5 HI ALARM
XRELAY 15
STRM 5 - 3
STREAM 5 HI-HI ALARM
XRELAY 16
STRM 6 - 1
STREAM 6 ACTIVE
XRELAY 17
STRM 6 – 2
STREAM 6 HI ALARM
XRELAY 18
STRM 6 - 3
STREAM 6 HI-HI ALARM
XVALVE 1
N/A
STREAM 1 INPUT VALVE
XVALVE 2
N/A
STREAM 2 INPUT VALVE
XVALVE 3
N/A
STREAM 3 INPUT VALVE
XVALVE 4
N/A
STREAM 4 INPUT VALVE
XVALVE 5
N/A
STREAM 5 INPUT VALVE
XVALVE 6
N/A
STREAM 6 INPUT VALVE
7.2.3.3.
Memory Reset
This function is currently disabled.
7.2.3.4.
Zero Calibration
The zero calibration allows the instrument to calibrate its internal ozone offset
factor. This should only be done with a source of zero air connected to the “Sample
In” port of the 465L. Allow the instrument to stabilize on the zero air source before
attempting to zero. This normally takes 10-15 minutes.
When entering the Zero Calibration menu, the prompt “ENT TO CAL” appears on
the display. Simply confirm the calibration by pressing the
ENT
button to perform
the calibration (to abort the calibration, press and hold the
CFG
button to return to
the previous menu level.) After pressing
ENT
, the instrument will automatically
exit the menu mode and return to Concentration mode. The concentration reading
should quickly go to zero. Please note that while measuring zero air, a certain
amount of noise or “dithering” of the concentration about the zero point will occur
and is normal. This noise is typically 1-3 ppb (0.001 - 0.003 ppm) in magnitude.