Trace Oxygen Analyzer
Operation
Teledyne
Analytical
Instruments
43
Readjust the gas pressure into the analyzer until the flowrate (as
read on the analyzer’s SLPM flowmeter) settles between 0.15 and 2.4
SLPM (approximately 0.2 - 5 SCFH).
If you are using password protection, you will need to enter your
password to gain access to either of these functions. Follow the
instructions in Sections 4.3.3 to enter your password. Once you have
gained clearance to proceed, you can enter the
Zero
or
Span
function.
4.8.1 Zero Cal
The
Zero
button on the front panel is used to enter the zero
calibration function. Zero calibration can be performed in either the
automatic or manual mode. In the
automatic
mode, an internal
algorithm compares consecutive readings from the sensor to determine
when the output is within the acceptable range for zero. In the
manual
mode, the operator determines when the reading is within the acceptable
range for zero. Make sure the zero gas is connected to the instrument. If
you get a
CELL FAILURE
message skip to Section 4.8.1.3.
4.8.1.1
A
UTO
M
ODE
Z
EROING
Press
Zero
to enter the zero function mode. The screen allows you
to select whether the zero calibration is to be performed automatically or
manually. Use the
▲
/
▼
arrow keys to toggle between
AUTO
and
MAN
zero settling. Stop when
AUTO
appears, blinking, on the display.
Zero: Settling: AUTO
<ENT> To Begin
Press
Enter
to begin zeroing.
#### PPM
Zero
Slope=#### ppm/s
The beginning zero level is shown in the upper left corner of the
display. As the zero reading settles, the screen displays and updates
information on
Slope
(unless the Slope starts within the acceptable zero
range and does not need to settle further).