Installation Model
3020T
Teledyne
Analytical
Instruments
30
obtainable voltage, depending on the load impedance applied. See
Figure 3-8.
Figure 3-8: FET Series Resistance
3.4 Installing the Micro-Fuel Cell
The Micro-Fuel Cell is not installed in the cell block when the
instrument is shipped. Install it before the analyzer is placed in service.
Once it is expended, or if the cell is exposed to air for too long, the
Micro-Fuel Cell will need to be replaced. The cell could also require
replacement if the instrument has been idle for too long.
When the micro-Fuel Cell needs to be installed or replaced, follow
the procedures in Chapter 5,
Maintenance,
for removing and installing
cells.
3.5 Gas Connections
Before using this instrument, it should be determined if the unit will
be used for pressurized service or vacuum service and low pressure
applications. Inspect the restrictor kit that came with the unit. The kit
consists of two restrictors and a union for 1/4” diameter tubing. Notice
that the two 1-3/4” long, 1/4” diameter tubing are restrictors. It has an
open end and a closed end with a small circular orifice. The restrictor
without the blue sticker is for low pressure and vacuum service. For high
pressure applications (5-50psig), use the restrictor that has a blue sticker
on the body.
For pressurized service (> 5psig) , use the restrictor
with
the blue
dot and union from the restrictor kit and attach it to the Sample In port.
The small circular orifice should face away from the back of the unit
(against the direction of gas flow).