Trace Oxygen Analyzer
Installation
Teledyne
Analytical
Instruments
25
Figure 3-6: Types of Relay Contacts
The connectors are:
Threshold Alarm 1:
Can be configured as high (actuates when concentration is above
threshold), or low (actuates when concentration is below
threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).
Threshold Alarm 2:
Can be configured as high (actuates when concentration is above
threshold), or low (actuates when concentration is below
threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).
System Alarm:
Actuates when DC power supplied to circuits is unacceptable in
one or more parameters.
Permanently configured as failsafe and latching. Cannot be
defeated.
Actuates if self test fails.