
4 Operation
Model 2010A
4-18 Part I
Teledyne Analytical Instruments
Use the
ΔΔΔΔΔ∇
keys to choose between % or ppm units. Then press
Enter
to
move to the next screen.
AL1: 1ØØØ ppm HI
Dft:N Fs:N Ltch:N
Five parameters can be changed on this screen:
•
Value of the alarm setpoint, AL1: ####
•
Out-of-range direction, HI or LO
•
Defeated? Dft:
Y/N
(Yes/No)
•
Failsafe? Fs:
Y/N
(Yes/No)
•
Latching? Ltch:
Y/N
(Yes/No).
•
To define the setpoint, use the
< >
arrow keys to move the
blinking over to AL1: ####. Then use the
Δ∇
arrow keys to
change the number. Holding down the key speeds up the
incrementing or decrementing.
•
To set the other parameters use the
< >
arrow keys to move the
blinking over to the desired parameter. Then use the
Δ∇
arrow
keys to change the parameter.
•
Once the parameters for alarm 1 have been set, press
Alarms
again, and repeat this procedure for alarm 2 (AL2).
•
To reset a latched alarm, go to
Dft–
and then press either
Δ
two
times or
∇
two times. (Toggle it to
Y
and then back to
N.
)
–OR –
Go to
Ltch–
and then press either
Δ
two times or
∇
two times.
(Toggle it to
N
and back to
Y.
)
4.6
The Range Select Function
The
Range
function allows you to manually select the concentration range
of analysis (
MANUAL
), or to select automatic range switching (
AUTO
).
In the
MANUAL
screen, you are further allowed to define the high and low
(concentration) limits of each Range, and select a single, fixed range to run.
CAUTION: If this is a linearized application, the new range must
be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.
Содержание 2010A
Страница 16: ...Teledyne Analytical Instruments 1 Part 1 Introduction Model 2010A ...
Страница 22: ...2 6 Part I 2 Operational Theory Model 2010A Teledyne Analytical Instruments ...
Страница 34: ...Teledyne Analytical Instruments 3 Installation Model 2010A 3 12 Part I ...
Страница 70: ...5 Maintenance Model 2010A 5 6 Part I Teledyne Analytical Instruments ...
Страница 86: ...1 4 Part II Teledyne Analytical Instruments 1 Introduction Model 2010A ...
Страница 104: ...4 Maintenance Model 2010A 4 8 Part II Teledyne Analytical Instruments ...