4 Operation
Model 2010A
4-10 Part I
Teledyne Analytical Instruments
RUNNING DIAGNOSTIC
Testing Preamp — Cell
When the testing is complete, the results are displayed.
Power: OK Analog: OK
Cell: 2 Preamp: 3
The module is functioning properly if it is followed by
OK
. A number
indicates a problem in a specific area of the instrument. Refer to Chapter
5
Maintenance and Troubleshooting
for number-code information. The results
screen alternates for a time with:
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.
4.3.6 The Model Screen
Move the < > arrow key to
MORE
and press
Enter
. With
MODEL
blinking, press
Enter
. The screen displays the manufacturer, model, and soft-
ware version information.
4.3.7 Checking Linearity with ALGORITHM
From the
System
Function screen, select
ALGORITHM
, and press
Enter
.
sel rng to set algo:
—> Ø1 Ø2 Ø3 <—
Use the < > keys to select the range: 01, 02, or 03. Then press
Enter
.
Gas Use: H2 — N2
Range: Ø — 10%
Press Enter again.
Algorithm setup:
VERIFY SET UP
Select and
Enter
VERIFY
to check whether the linearization has been
accomplished satisfactorily.
Dpt INPUT OUTPUT
Ø Ø.ØØ Ø.ØØ
Содержание 2010A
Страница 16: ...Teledyne Analytical Instruments 1 Part 1 Introduction Model 2010A ...
Страница 22: ...2 6 Part I 2 Operational Theory Model 2010A Teledyne Analytical Instruments ...
Страница 34: ...Teledyne Analytical Instruments 3 Installation Model 2010A 3 12 Part I ...
Страница 70: ...5 Maintenance Model 2010A 5 6 Part I Teledyne Analytical Instruments ...
Страница 86: ...1 4 Part II Teledyne Analytical Instruments 1 Introduction Model 2010A ...
Страница 104: ...4 Maintenance Model 2010A 4 8 Part II Teledyne Analytical Instruments ...