Appendix A – Liquid Cooling Setup Instructions
Warning: Use of equipment not originally provided by Teledyne Photometrics for use with
liquid cooled cameras will void any and all warranty coverage of the product.
1.
Unpack the cooler and hose assembly.
2.
Confirm the cooler and hoses are pre-filled with blue-colored coolant.
3.
Press one hose connector into its mate on the cooler.
4.
Repeat with the second hose connector.
5.
Press each hose connector onto its mating connector on the camera:
•
Circulator Out > Camera In
•
Camera Out > Circulator In
•
Listen for the “click.”
6.
Pull each connector to ensure they are locked.
7.
Inspect the setup to ensure the hose connectors are secure at the cooler and camera.
8.
Set both the pump speed and fan speed to level 10 on the front display on the
circulator.
9.
Plug in the cooler and turn it on.
10.
Look through the clear cover on the coolant reservoir to observe the liquid level and
confirm circulation. (The liquid surface will appear agitated with normal circulation.)
11.
Turn on the camera power and continue the setup per the
QuickStart Guide.
12.
Use software to set the camera fan speed to “off."
DO NOT turn off camera fan
without liquid circulating.
Please note there is no temperature adjustment on the unit.