OLCT 80
GAS TRANSMITTER
USER MANUAL
108
NPO80GB
Revision F.1
Cause of the error or fault
Display
Corrective action
OLCT 80
temperature out of
range.
Protect the
OLCT 80
from extreme variations in
temperature. See
Operating temperature
under
Metrological characteristics
on page 92.
Line voltage too low
Check the voltage of the
OLCT 80
's power supply.
See
Power supply at detector terminals
under
Metrological characteristics
on page 92.
Other fault codes that may be displayed include the following:
•
eeprom
•
def ana1
•
def ana2
•
def ir elc
•
def ir opt
•
def RAMint
•
def RAMext
•
def ROM
•
def eeprom
•
MEM perdue
These faults require a maintenance technician and cannot be resolved by the user.
Содержание OLCT 80 d
Страница 1: ...NPO80GB Revision F 1 OLCT 80 GAS TRANSMITTER USER MANUAL...
Страница 10: ...OLCT 80 GAS TRANSMITTER USER MANUAL 4 NPO80GB Revision F 1...
Страница 56: ...OLCT 80 GAS TRANSMITTER USER MANUAL 50 NPO80GB Revision F 1...
Страница 78: ...OLCT 80 GAS TRANSMITTER USER MANUAL 72 NPO80GB Revision F 1...
Страница 80: ...OLCT 80 GAS TRANSMITTER USER MANUAL 74 NPO80GB Revision F 1...
Страница 86: ...OLCT 80 GAS TRANSMITTER USER MANUAL 80 NPO80GB Revision F 1...
Страница 92: ...OLCT 80 GAS TRANSMITTER USER MANUAL 86 NPO80GB Revision F 1 12 1 OLCT 80 without and with antenna...
Страница 93: ...OLCT 80 GAS TRANSMITTER USER MANUAL NPO80GB Revision F 1 87...
Страница 94: ...OLCT 80 GAS TRANSMITTER USER MANUAL 88 NPO80GB Revision F 1...
Страница 95: ...OLCT 80 GAS TRANSMITTER USER MANUAL NPO80GB Revision F 1 89 12 2 IR20 remote control of the OLCT 80...
Страница 96: ...OLCT 80 GAS TRANSMITTER USER MANUAL 90 NPO80GB Revision F 1...
Страница 104: ...OLCT 80 GAS TRANSMITTER USER MANUAL 98 NPO80GB Revision F 1...
Страница 115: ...OLCT 80 GAS TRANSMITTER USER MANUAL NPO80GB Revision F 1 109...