![Teledyne Lecroy PCI Express 5.0 M.2 M-Key Interposer Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/teledyne-lecroy/pci-express-5-0-m-2-m-key-interposer/pci-express-5-0-m-2-m-key-interposer_user-manual-and-quick-start-manual_1081011005.webp)
Hardware Installation
To use this interposer:
1. Set the SW2 DIP switches to the desired clock selection.
Note:
“HOST_CLK” should be the default setting unless you are using an external clock reference source.
2. Move the stand off from the default position to the position required by the DUT.
3. Install the M.2 SSD device under test (DUT) into the connector on the interposer as shown (will fit 30 mm, 42mm,
60mm, 80mm and 110mm).
4. Connect the Summit Analyzer to the interposer using the high speed serial cables (If using Summit T54 use Gen5 Y
cable PE028UCA-X. If using Summit T54 with Gen5 Phy use Gen5 straight cable PE027UCA-X).
5. Connect the analyzer to a host machine using the USB port on the front panel of the Summit analyzer.
6. If not already done, install the PCIe Protocol Analysis on the host machine.
7. Connect 12V DC using the AC adapter supplied with the interposer. (Make sure that the AC adapter is powered on).
8. Power on the Analyzer and wait for it to be recognized by the application.
Note
: If prompted to update the firmware
please do that before proceeding.
9. Power on the host machine.
10. Launch the Teledyne LeCroy software application to monitor, record and view PCI Express traffic passing through the
M.2 Interposer.
11. Power on the system under test.
Note:
Step 8, 9, and 10 are needed in this order for “Power On
”
traces.
Note:
Use the link width control button SW1 to match the minimum physical lane width of the host and device.
PCI Express Gen5 M.2 MKey Interposer CrossSync Phy Interposer Interconnection Overview
Keyed paddle board
inserts in M.2 connector
12 V DC from power supply
Upstream connection to oscilloscope
Reference clock probe connections
Summit T54 Gen5 Phy
Power rail probe connections
External clock inputs
Device under test (DUT) Compatible with
30x30, 30x42, 30x60, 30x80, 30x110 and
22x30, 22x42, 22x60, 22x80, 22x110
SSD carrier board
Sideband signal connections
Downstream connection to oscilloscope
PE028UCA-X Y Cable (Use with T54 Expanded Mode or Summit T516)
PE027UCA-X straight cable (Use with Summit T54 or Summit T516 MultiPort Mode)
5