CP03x/CP03xA Current Probes
Description
Minimum Requirements
Example Equipment
Wide Band Oscilloscope
200 MHz bandwidth
ProBus interface equipped
Teledyne LeCroy WaveRunner 6 Zi
Digital Multimeter (2)
DC: 0.1% Accuracy
5½ digit resolution
Agilent Technologies 34401A
Fluke 8842A-09
Function Generator
50 Hz sine wave output 3 Vrms
into 50Ω
Teledyne LeCroy WaveStation 3082
Stanford Research Model DS340
Calibration Fixture
100 t
urn loop in series with 0.5Ω
±0.1% resistor with sense
terminals
Teledyne LeCroy CP031-CF02
Calibration Fixture
ProBus Extension Cable
Teledyne LeCroy PROBUS-CF01
Banana Plug Adapter
Female BNC to Dual Banana Plug Pomona 1269
Patch Cables (4)
Male Banana to Male Banana, 12" Pomona B-12-0 (black)
Pomona B-12-2 (red)
BNC Adapter
BNC Male to Dual Banana Jack
Pomona 1296
Preliminary Procedure
1.
Connect the probe to the C1 input of the instrument and completely close the
probe slider.
2.
Turn on the instrument and allow at least 30 minutes warmup time for the probe
before performing the verification procedure.
3.
Turn on the other test equipment and allow these to warm up for the time
recommended by the manufacturer.
4.
While the instruments are reaching operating temperature, make a photocopy of
the Performance Verification Test Record and fill in the necessary data.
12