Operator’s Manual
3.
Close the slider on the sensor head until the "UNLOCK’ indication disappears
and the "LOCK" indication appears.
4.
Verify that the opening lever is firmly locked and the clamp is securely closed.
CAUTION:
Never use this probe on bare conductors. The core and shield are
grounded and any voltage applied to the conductor may cause damage the
probe or the circuit under test.
Operating with an Oscilloscope
When the probe is connected to a Teledyne LeCroy oscilloscope, the displayed scale
factor and measurement values will be adjusted to account for the effective gain of the
probe.
A/div, Offset, and Coupling, and Bandwidth (BWL) are controlled from the Channel
setup (Cx) dialog.
Channel setup dialog with Probe dialog behind it.
The probe’s attributes are shown on the Probe dialog, which appears behind the
Channel dialog when a probe is detected. The Probe dialog also contains controls for
Auto Zero and Degauss.
Probe dialog.
9