09421A DCN8343
Teledyne API Model 480H O
3
Monitor User Manual
33
VAR
DESCRIPTION
Default Value
Range
REF GAIN
Gain applied for zero calibration.
1
FAILSAFE OPERATION
Relays and digital outputs operate in
fail safe operation. Outputs are
energized in normal operation and de-
energized in fault
OFF
ON/OFF
AZERO INTERVAL
Interval at which zero calibration is
performed. (minutes)
6
3-1440
AZERO ENABLE
Enables/Disengages zero calibration
ON
ON/OFF
ANALOG_RANGE
Full-scale concentration range. Used
to scale the analog output.
20
5-50 g/Nm3
0.5-25.0
WT%
ANALOG HW CONF
Determines if the analog output uses
4-20mA or 0-5V output
4-20mA
4-20mA, 0-5V
O3 OFFSET
[Not available in this model]
O3 SLOPE
Calibration slope value
1.0
0.5 - 1.5
USER SLOPE
User-editable calibration slope value
1.0
0.5 – 1.5
USER UNITS
Setting for concentration unit-of-
measure.
WT%, g/Nm3 (user selectable)
CARRIER GAS WEIGHT
Molecular weight of carrier gas (i.e.,
feed gas for ozone generator). This
value is only used for the calculation of
wt% concentration units.
32 g/mol
Содержание Everywhereyoulook 480H
Страница 2: ...E IS INTENTIONALLY LEFT BLANK...
Страница 6: ...iv Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 This page intentionally left blank...
Страница 12: ...10 Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 This page intentionally left blank...
Страница 17: ...09421A DCN8343 Teledyne API Model 480H O3 Monitor User Manual 15...
Страница 40: ...38 Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 Figure 6 1 RS 232 Signal Configurations...