30
Teledyne API Model 480H O
3
Monitor User Manual
09421A DCN8343
5.2.2. SENSOR MENU
The Sensor menu allows the operator to view various measurement parameters in
real-time. This can be useful for diagnosing various instrument or system problems.
Note that these values are updated in real-time as they are repeatedly scrolled on the
display.
Table 5-2. Sensor Menu Parameters
Parameter
Description
Units
Normal
Range*
MEAS
UV detector reading, measure cycle
mV
250 – 3120
REF
UV detector reading, reference cycle
mV
250 – 3120
CELL TEMP
Measurement of cell temperature,
depends on ambient conditions
°C
5 - 50
O3 CELL PRESSURE
Sample pressure
psia
9.0 – 18.0
FLOW
Sample Gas Flow Rate
cc/min 340 – 1860
*After initial 5 minute warm-up period.
5.2.3. DIAG MENU
The DIAG menu contains functions that are useful for testing and configuring
external equipment that may be connected to the rear panel I/O of the instrument.
5.2.3.1. A
NALOG
S
TEP
This menu exercises the analog output of the system through a five-point step test
that corresponds to the ozone concentration range. With this function, the
instrument will automatically step the analog output (see Section 3.4.1 for details on
the analog output connection) through 5 points from 0 to 100 %, representing the
output configuration of either Voltage or Current (see Table 5-3 for values). The
operator can manually control the stepping by pressing either one of the
UP
/
DOWN
arrow buttons.
Table 5-3. Analog Step Test Values
Point
0-5V Output
4-20 mA Output
0%
0.00 V
4 mA
25%
1.25 V
8 mA
50%
2.50 V
12 mA
75%
3.75 V
16 mA
100%
5.00 V
20 mA
Press the Back button to exit from the Analog Step Test function.
Содержание Everywhereyoulook 480H
Страница 2: ...E IS INTENTIONALLY LEFT BLANK...
Страница 6: ...iv Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 This page intentionally left blank...
Страница 12: ...10 Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 This page intentionally left blank...
Страница 17: ...09421A DCN8343 Teledyne API Model 480H O3 Monitor User Manual 15...
Страница 40: ...38 Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 Figure 6 1 RS 232 Signal Configurations...