20
Teledyne API Model 480H O
3
Monitor User Manual
09421A DCN8343
3.4.1.1. A
NALOG
O
UTPUT
H
ARDWARE
V
ERIFICATION OR
C
ONVERSION
To change or verify the configuration of the analog output, refer to Figure 3-5 while
performing the following instructions:
1.
Disconnect power from the monitor.
2.
Open front panel.
3.
Locate J15 and J17 connectors on the Mainboard PCA (remove Relay Expansion
board if installed as indicated in Figure 3-5).
4.
Set the desired output as shown in Figure 3-5 where:
•
for Current output (4-20 mA), jumper Pins 1-2 on both J15 and J17, or
•
for Voltage output (0-5 VDC, jumper Pins 2-3 on both J15 and J17.
5.
If applicable, reinstall/connect Relay Expansion board.
6.
Re-secure the front panel (NEMA Configuration).
7.
Reconnect power to the monitor.