Portable Trace Oxygen Analyzer
Appendix
Teledyne Analytical Instruments
33
Appendix
A.1 Specifications
Certification:
ISSeP08ATEX036X
II 1 G
Ex ia IIC Tx
CE xxxx
Standards:
EN 60079-0: 2006 (IEC 60079-0: 2004)
EN 60079-11: 2007 (IEC 60079-11: 2006)
EN 60079-26: 2004 (IEC 60079-26: 2004)
EN 60079-0: 2012 (IEC 60079-0: 2011)
EN 60079-11: 2012 (IEC 60079-11: 2011)
EN 60079-26: 2007 (IEC 60079-26: 2006)
Micro-fuel Cell Class:
B-2CXL, A-2CXL
Ranges of Analysis:
1 user defined range
0-2 ppm to 0-25%
Accuracy:
±2% of scale or ±1 PPM (whichever is
greater) at constant temperature;
±5% of reading or ±1 PPM (whichever is
greater) over the operating temperature
range.
Recommended Span Gas:
Atmospheric air for percent analysis or
80-90% of range most likely to be used
Signal Output:
0-1 VDC
Electrical safety parameters:
Uo = 4.65 V, Io = 193.74 mA,
Po = 225.2 mW, Co = 90 uF,
Lo = 0.7 mH, Lo/Ro = 157.86 uH/Ohm
Dimensions:
10.9” x 6.2” x 5.2” deep
Weight:
4.4 lbs. (2.0 kg.)
Power Requirements:
100-240 VAC 47-63 Hz Charging Adapter