Operation
Model 3110XL
Teledyne Analytical Instruments
20
Concentration
(ppm)
Output
(V)
Concentration
(ppm)
Output
(V)
0
0
60
0.6
10
0.1
70
0.7
20
0.2
80
0.8
30
0.3
90
0.9
40
0.4
100
1.0
50
0.5
Use the UP/DOWN keys to navigate to the
AO RNG
screen and
then press ENTER. The following screen appears:
To define or change the range setting:
1.
Use the UP /DOWN keys to change the range value. Press
ENTER when the desired value is shown. The value will be
accepted and the cursor will move to the next editable field
(units).
2.
Use the UP/DOWN keys to cycle between ppm and percent
for units. Press ENTER to accept the displayed unit. The
display will then return to the
HOME
screen.
Note: The range setting resolution is 0.1 ppm on a range set to
less that 1% or 0.01 % for percent analysis.
3.2.6 SPAN VALUE Screen
The
SPAN VALUE
screen displays the oxygen concentration of the
span gas used for calibration. This is not a measured value; it is the
known span gas concentration that is input to the analyzer by the
operator.
Содержание 3110XL
Страница 10: ......
Страница 16: ...Introduction Model 3110XL Teledyne Analytical Instruments 6 BLANK PAGE...
Страница 24: ...Installation Model 3110XL Teledyne Analytical Instruments 14...
Страница 36: ...Operation Model 3110XL Teledyne Analytical Instruments 26...