Paramagnetic Oxygen Analyzer
Operation
Teledyne Analytical Instruments
49
If you are using password protection, you will need to enter your
password to gain access to either of these functions. Follow the
instructions in Sections 4.3.3 to enter your password. Once you have
gained clearance to proceed, you can enter the
Zero
or
Span
function.
4.8.1 Zero Cal
The
Zero
button on the front panel is used to enter the zero
calibration function. Zero calibration can be performed in either the
automatic or manual mode. In the
automatic
mode, an internal
algorithm compares consecutive readings from the sensor to determine
when the output is within the acceptable range for zero. In the
manual
mode, the operator determines when the reading is within the acceptable
range for zero. Make sure the zero gas is connected to the instrument.
4.8.1.1
A
UTO
M
ODE
Z
EROING
Press
Zero
to enter the zero function mode. The screen allows you
to select whether the zero calibration is to be performed automatically or
manually. Use the UP/DOWN switch to toggle between
AUTO
and
MAN
zero settling. Stop when
AUTO
appears, blinking, on the display.
Zero: Settling: AUTO
<ENT> To Begin
Select
Enter
to begin zeroing.
####
PPM Zero
Slope=####
ppm/s
The beginning zero level is shown in the upper left corner of the
display. As the zero reading settles, the screen displays and updates
information on
Slope
(unless the Slope starts within the acceptable zero
range and does not need to settle further).
Then, and whenever Slope is less than 0.08 for at least 3 minutes,
instead of Slope you will see a countdown:
5 Left
,
4 Left
, and so forth.
These are five steps in the zeroing process that the system must
complete, AFTER settling, before it can go back to
Analyze
.
Содержание 3020 M
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 11: ...Model 3020M ATEX Teledyne Analytical Instruments iv Blank Page...
Страница 17: ...Model 3020M ATEX Teledyne Analytical Instruments vi Blank Page...
Страница 33: ...Introduction Model 3020M ATEX Teledyne Analytical Instruments 10 Blank Page...
Страница 43: ...Operational Theory Model 3020M ATEX Teledyne Analytical Instruments 20 Blank Page...
Страница 61: ...Installation Model 3020M ATEX Teledyne Analytical Instruments 38 Blank Page...
Страница 90: ......
Страница 92: ......
Страница 94: ......
Страница 96: ......
Страница 98: ......
Страница 100: ......
Страница 102: ......
Страница 104: ......
Страница 105: ...C TELEDYNE...
Страница 106: ......
Страница 107: ......