Installation
Model 3020M-ATEX
Teledyne Analytical Instruments
28
To provide an indication of the range, a second pair of analog
output terminals are used. They generate a steady preset voltage (or
current when using the current outputs) to represent a particular range.
Table 3-2 gives the range ID output for each analysis range.
Table 3-2: Analog Range ID Output—Example
Range
Voltage (V)
Current (mA)
LO
0.25
8
MED
0.50
12
HI
0.75
16
CAL (0-25%)
1.00
20
3.3.4 Alarm Relays
There are three alarm-circuit connectors on the alarm relays block
(under RELAY OUTPUTS) for making connections to internal alarm
relay contacts. Each provides a set of Form C contacts for each type of
alarm. Each has both normally open and normally closed contact
connections. The contact connections are indicated by diagrams on the
connector panel. They are capable of switching up to 3 amperes at
250 V ac into a resistive load. See Figure 3-8.
Figure 3-8: Types of Relay Contacts
Содержание 3020 M
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 11: ...Model 3020M ATEX Teledyne Analytical Instruments iv Blank Page...
Страница 17: ...Model 3020M ATEX Teledyne Analytical Instruments vi Blank Page...
Страница 33: ...Introduction Model 3020M ATEX Teledyne Analytical Instruments 10 Blank Page...
Страница 43: ...Operational Theory Model 3020M ATEX Teledyne Analytical Instruments 20 Blank Page...
Страница 61: ...Installation Model 3020M ATEX Teledyne Analytical Instruments 38 Blank Page...
Страница 90: ......
Страница 92: ......
Страница 94: ......
Страница 96: ......
Страница 98: ......
Страница 100: ......
Страница 102: ......
Страница 104: ......
Страница 105: ...C TELEDYNE...
Страница 106: ......
Страница 107: ......