General Maintenance
4- 8
WFM6100, WFM7000, and WFM7100 With Option FP Service Manual
Table 4- 4: Symptoms and causes (Cont.)
Symptom
Possible sources or recommended detailed troubleshooting procedure to follow
Fails any of These Power On Self Tests
(POST)
H
Eye Hardware Revision
H
Eye Board Cable Version:
H
Eye Register Readback:
H
Eye Board Power Supply
H
Eye Board Cal EEPROM
Review messages in diagnostic log.
Perform primary power supply checks.
Perform secondary power supply checks.
Run Advanced Diagnostics and look for other information.
Replace the EYE/PHY board.
Diagnostic log shows Fail on any of these
Advanced Diagnostics:
ADV_DIAG_LSS
Review messages in diagnostic log. If the waveform monitor also fails other tests as well,
then suspect the Main board and perform the primary and secondary power supply tests.
If only this test fails, then perform
Isolating Advanced Diagnostic Lissajous Errors
procedure to isolate the problem to Main or Audio board.
Содержание WFM6100 Opt. MB
Страница 4: ......
Страница 12: ...Service Safety Summary viii WFM6100 WFM7000 and WFM7100 With Option FP Service Manual ...
Страница 64: ...Repackaging Instructions 4 34 WFM6100 WFM7000 and WFM7100 With Option FP Service Manual ...
Страница 76: ...Replaceable Parts 5 12 WFM6100 WFM7000 and WFM7100 With Option FP Service Manual ...