Theory of Operation
2- 6
WFM6100, WFM7000, and WFM7100 With Option FP Service Manual
come from any input. The selected signal output from the FPGA is converted
from digital to analog, and then sent as a balanced output to the 62-pin rear panel
analog audio connector.
The AES B connectors can be configured as outputs for use with any input type.
They can serve as an active loopthrough if the input is AES, a de--embedder if
the input is embedded, or an A/D converter if the input is analog. One decoded
Dolby audio pair may be output if Option DD is installed; up to five decoded
Dolby audio pairs may be output if Option DDE is installed.
Option EYE/PHY
Option EYE is for basic SDI signal monitoring, providing Eye pattern displays
and readouts of jitter and key cable parameters. Option PHY has these same
capabilities, plus jitter waveform display and automated Eye parameter
measurements. Options EYE and PHY provide the following features:
H
Eye pattern display
H
Equalized Eye pattern display
H
Jitter readout
H
Jitter Meter
H
Cable Loss readout
H
Approx Cable readout
H
Source Level readout
H
Jitter waveform display
1
H
Eye Amplitude readout
1
H
Eye Risetime readout
1
H
Eye Falltime readout
1
H
Eye Rise--Fall (difference) readout
1
1
Requires Option PHY.
Options EYE and PHY support both HD and SD SDI signals, limited by the
capability of the instrument they are installed in:
H
WFM6100: SD only.
H
WFM7100: HD only, SD only, or both, depending on installed options.