Adjustments
3- 4
WFM6100, WFM7000, and WFM7100 With Option FP Service Manual
11.
Use the waveform monitor’s General knob to adjust the signal amplitude
(not including the sync pulse) to 700 mV on the oscilloscope.
12.
Follow the instructions at the bottom of the screen to Save and Exit
Calibration mode.
1.
Connect a 100 IRE Flat Field video signal to the
Cmpst A
input. Terminate
the loop-through in 75
Ω
.
2.
Press
Input C
to select the Composite A signal.
3.
Press the
CONFIG
button.
4.
Use the arrow keys or the General knob to navigate to
Outputs
and set the
Composite Pixmon
to
On
.
5.
Press the left arrow key to return to the first level of the CONFIG menu.
6.
Connect an XGA to 5x BNC adapter cable to the PixMon output on the
waveform monitor.
7.
Connect the “Green” video signal to the oscilloscope input. Use an oscillo-
scope with a 75
Ω
input, or use a 75
Ω
feed through termination.
8.
Set the Oscilloscope:
Vert
200 mV/Div
Horiz
20
s/Div
Trigger
Edge
9.
Press the
Config
button.
10.
Select
Utilities
>
Calibration
, and then press
SEL
to enter the calibration
menu.
11.
Navigate to the
Pixmon Composite Gain
case, and then press
SEL
.
12.
Adjust the General Purpose knob to set the video portion of the signal to
714 mV (100 IRE).
13.
Follow the instructions at the bottom of the screen to Save and Exit
Calibration mode.
1.
Perform the following steps if the Offset is outside the specification:
2.
Press the
Config
button.
3.
Select
Utilities
>
Calibration
, and then press
SEL
to enter the calibration
menu.
PixMon (Composite)
Composite Analog Input
DC Offset Restore Off
(Option CPS)