Speci
fi
cations
Warranted Speci
fi
cations (cont.)
Characteristic
Description
TLA5000B
Maximum window size = Maximum channel-to-channel skew + (2 x sample
uncertainty) + system var 100 ps
TLA7ACx (single module)
Maximum window size = Maximum channel-to-channel skew + (2 x sample
uncertainty) + 100 ps Maximum setup time = User interface setup time + 75 ps
Maximum hold time = User interface hold time + 50 ps
TLA7Bxx
220 ps maximum (180 ps Typical) single channel on single module
240 ps maximum (200 ps Typical) single channel on merged module
Maximum synchronous clock rate
TLA7N4
200 MHz in full-speed mode (5 ns minimum between active clock edges)
3
100 MHz in half-speed mode (10 ns minimum between active clock edges)
3
TLA5000B
235 MHz in full speed mode (4.25 ns minimum between active clock edges)
TLA7ACx
120 MHz in quarter-speed mode (8.3 ns minimum between active clock edges)
3
235 MHz in half-speed mode (4.25 ns minimum between active clock edges)
3
450 MHz in full-speed mode (2.2 ns minimum between active clock edges)
800 MHz on half channels
4
Software controls the selection between full-speed and half-speed modes.
TLA7Bxx
750 MHz, one sample point per clock, all channels
750 MHz, four samples points per clock, half channels
1400 MHz, one sample point per clock, all channels
1400 MHz, two sample points per clock, half channels
Trigger System
Trigger State sequence rate
Same rate as valid data samples received; 500 MHz maximum (TLA7Bxx,
800 MHz maximum)
1
It is possible to use storage control and only store data when it has changed (transitional storage).
2
Applies to asynchronous sampling only. Setup and hold window speci
fi
cation applies to synchronous sampling only.
3
Software controls the selection between full-speed, half-speed, and quarter-speed modes.
4
This is a special mode and has some limitations such as the clocking state machine and trigger state machine only running at 500 MHz.
TLA Quick Start User Manual
45