![Tektronix TLA5201B Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/tektronix/tla5201b/tla5201b_quick-start-user-manual_1078211051.webp)
Speci
fi
cations
Speci
fi
cations
This section provides the warranted and mechanical speci
fi
cations for the TLA5000B Series Logic Analyzers, TLA7000
Series Logic Analyzers, TLA7Bxx, TLA7ACx, and TLA7N4 Series Logic Analyzer. All speci
fi
cations (marked with the.
symbol) are guaranteed and can be checked directly or indirectly at your nearest Tektronix location or by following the
procedures described in individual service manuals or performance veri
fi
cation documents. Typical speci
fi
cations are
provided for your convenience but are not guaranteed.
The performance limits in these speci
fi
cations are valid under the following conditions:
The instrument must be in an environment with temperature, altitude, humidity, and vibration within the operating limits
described in these speci
fi
cations.
Logic analyzer modules must be installed in a logic analyzer mainframe.
The instrument must have had a warm-up period of at least 30 minutes.
The instrument must have been calibrated and adjusted at an ambient temperature b20 °C and +30 °C.
Atmospheric characteristics
Characteristic
Description
Temperature:
Operating and nonoperating
Operating (no media in
fl
oppy disk drive or DVD drive):
+5 °C to +50 °C, 15 °C/hr maximum gradient, noncondensing (derated 1 °C
per 305 m (1000 ft) above 1524 m (5000 ft) altitude)
1 2
Nonoperating (no media in
fl
oppy disk drive or DVD drive):
–20 °C to +60 °C, 15 °C/hr maximum gradient, noncondensing
Relative Humidity:
Operating and nonoperating
Operating (no media in
fl
oppy disk drive or DVD drive):
20% to 80% relative humidity, noncondensing. Maximum wet bulb temperature:
+29 °C (derates relative humidity to approximately 22% at +50 °C)
3 4
Nonoperating (no media in
fl
oppy disk drive or DVD drive):
8% to 80% relative humidity, noncondensing. Maximum wet bulb temperature:
+29 °C (derates relative humidity to approximately 22% at +50 °C)
5
Altitude:
Operating and nonoperating
Operating:
To 3000 m (9843 ft), (derated 1 °C per 305 m (1000 ft) above 1524 m (5000 ft)
altitude
Nonoperating:
12190 m (40,000 ft)
1
TLA7Bxx and TLA7ACx series module operating temperature is +40 °C maximum.
2
TLA7012 operating temperature is 45 °C maximum due to the CD-RW DVD drive.
3
TLA7Bxx series module relative humidity derates to approximately 57% at +40 °C.
4
TLA7Bxx and TLA7ACx series module operating humidity is 5% to 90% up to +30 °C, 75% from +30 to +40 °C, noncondensing. Maximum
wet-bulb temperature is +29.4 °C.
5
TLA7Bxx and TLA7ACx series module nonoperating humidity is 5% to 90% limited by a wet bulb temperature of +40 °C.
42
TLA Quick Start User Manual