Command Descriptions
2–194
TDS Family Oscilloscope Programmer Manual
MEASUrement:REFLevel:PERCent:LOW
Sets or queries the percent, where 100% is equal to HIGH, that is used to
calculate the low reference level when MEASUrement:REFLevel:METHod is set
to PERCent. This command is equivalent to setting the Reference Levels in the
Measure menu.
Measurement
!
!
!
ranges from 0 to 100 percent, and is the low reference level. The default is
10%.
might return 15, meaning that the low reference level is 15% of HIGH.
MEASUrement:REFLevel:PERCent:MID
Sets or queries the percent, where 100% is equal to HIGH, that is used to
calculate the mid reference level when MEASUrement:REFLevel:METHod is
set to PERCent. This command is equivalent to setting the Reference Levels in
the Measure menu.
Measurement
!
!
Group
Syntax
Arguments
Examples
Group
Syntax