Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–95
*DDT?
<QString>
*DDT
<Space>
?
<Block>
<Block>
or
<QString>
is a complete sequence of program messages. The
messages can contain only valid commands that must be separated by semico-
lons and must follow all rules for concatenating commands (see page 2–4). The
sequence must be
≤
80 characters.
<Block>
format is always returned as a query
response.
*DDT #0ACQUIRE:STATE RUN;BELL<EOI>
specifies that the acquisition system will be started and the bell rings each time a
*TRG command is sent.
DELEte:SETUp (No Query Form)
Removes stored setups from memory and initializes the location with the factory
default setup.
NOTE
. The setup information cannot be recovered once it has been deleted.
Save and Recall
*RCL, RECAll:SETUp, *RST, *SAV, SAVe:SETUp, TEKSecure
DELEte:SETUp { <NR1> | ALL }
DELEte
:
SETUp
<Space>
ALL
<NR1>
<NR1>
is a value in the range 1 to 10, and specifies a setup storage location.
Using an out-of-range value causes an execution error.
ALL
specifies all the stored setups.
Arguments
Examples
Group
Related Commands
Syntax
Arguments
Содержание TDS 410A
Страница 4: ......
Страница 10: ...Table of Contents vi TDS Family Oscilloscope Programmer Manual ...
Страница 56: ...Command Groups 2 38 TDS Family Oscilloscope Programmer Manual ...
Страница 346: ...Status and Events 3 20 TDS Family Oscilloscope Programmer Manual ...
Страница 352: ...Program Examples 4 6 TDS Family Oscilloscope Programmer Manual ...
Страница 358: ...Appendix B Reserved Words B 4 TDS Family Oscilloscope Programmer Manual ...
Страница 382: ...Index Index 14 TDS Family Oscilloscope Programmer Manual ...