Application Examples
TDS 200-Series Digital Oscilloscope User Manual
45
Measuring Ring Frequency
To measure the ring frequency at the rising edge of a signal, do these
steps:
1. Push the CURSOR button to see the Cursor menu.
2. Push the top menu box button to select Time.
3. Use the CURSOR 1 knob to place a cursor on the first peak of
the ring.
4. Use the CURSOR 2 knob to place a cursor on the second peak of
the ring.
You can see the delta time and frequency (the measured ring
frequency) in the Cursor menu.