Basic Concepts
TDS 200-Series Digital Oscilloscope User Manual
19
The next table lists the time bases that you should use to avoid
aliasing at various frequencies and the respective sample rate.
Time base
Samples
per second
Maximum
frequency
Time base
Samples
per second
Maximum
frequency
1.0
m
s
250.0 MS/s
125.0 MHz*
5.0 ms
50.0 kS/s
25.0 kHz
2.5
m
s
100.0 MS/s
50.0 MHz*
10.0 ms
25.0 kS/s
12.5 kHz
5.0
m
s
50.0 MS/s
25.0 MHz*
25.0 ms
10.0 kS/s
5.0 kHz
10.0
m
s
25.0 MS/s
12.5 MHz*
50.0 ms
5.0 kS/s
2.5 kHz
25.0
m
s
10.0 MS/s
5.0 MHz*
100.0 ms
2.5 kS/s
1.25 kHz
50.0
m
s
5.0 MS/s
2.5 MHz
250.0 ms
1.0 kS/s
500.0 Hz
100.0
m
s
2.5 MS/s
1.25 MHz
500.0 ms
500.0 S/s
250.0 Hz
250.0
m
s
1.0 MS/s
500.0 kHz
1.0 s
250.0 S/s
125.0 Hz
500.0
m
s
500.0 kS/s
250.0 kHz
2.5 s
100.0 S/s
50.0 Hz
1.0 ms
250.0 kS/s
125.0 kHz
5.0 s
50.0 S/s
25.0 Hz
2.5 ms
100.0 kS/s
50.0 kHz
* Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
There are various ways to prevent aliasing: adjust the horizontal
scale, push the AUTOSET button, or change the acquisition mode.
NOTE
. If aliasing occurs, change to the Peak Detect acquisition
mode (described on page 16). This mode samples the highest and
lowest values so that the oscilloscope can detect faster signals.
Содержание TDS 200-Series
Страница 8: ...Table of Contents iv TDS 200 Series Digital Oscilloscope User Manual...
Страница 48: ...Operating Basics 36 TDS 200 Series Digital Oscilloscope User Manual...
Страница 118: ...Appendix B Accessories 106 TDS 200 Series Digital Oscilloscope User Manual...
Страница 120: ...General Care and Cleaning 108 TDS 200 Series Digital Oscilloscope User Manual...
Страница 126: ...Glossary 114 TDS 200 Series Digital Oscilloscope User Manual...
Страница 132: ...Index 120 TDS 200 Series Digital Oscilloscope User Manual...