![Tektronix TDS 200-Series Скачать руководство пользователя страница 109](http://html1.mh-extra.com/html/tektronix/tds-200-series/tds-200-series_user-manual_1079117109.webp)
Appendix A: Specifications
TDS 200-Series Digital Oscilloscope User Manual
97
Specifications (Cont.)
Trigger
Trigger Sensitivity,
Coupling
Sensitivity
Edge Trigger Type
DC
CH 1, CH 2,
CH 3 & CH 4
1 div from DC to 10 MHz*,
1.5 div from 10 MHz* to Full
TDS 210 and TDS 220 only
EXT
100 mV from DC to 10 MHz*,
150 mV from 10 MHz* to Full
EXT/5
500 mV from DC to 10 MHz*,
750 mV from 10 MHz* to Full
Trigger Sensitivity,
Coupling
Sensitivity
Edge Trigger Type,
AC
Same as DC at 50 Hz and above
typical
NOISE REJ
Reduces the DC-coupled trigger sensitivity by
2 times for > 10 mv/div to 5 V/div
HF REJ
Same as the DC-coupled limit from DC to
7 kHz, attenuates signals above 80 kHz
LF REJ
Same as the DC-coupled limits for frequencies
above 300 kHz, attenuates signals below
300 kHz
Trigger Level Range
Source
Range
Internal
±
8 divisions from center of screen
TDS 210 and TDS 220 only
EXT
±
1.6 V
EXT/5
±
8 V
* Bandwidth is not valid for the P2100 probe when the switch is set to 1X.
Содержание TDS 200-Series
Страница 8: ...Table of Contents iv TDS 200 Series Digital Oscilloscope User Manual...
Страница 48: ...Operating Basics 36 TDS 200 Series Digital Oscilloscope User Manual...
Страница 118: ...Appendix B Accessories 106 TDS 200 Series Digital Oscilloscope User Manual...
Страница 120: ...General Care and Cleaning 108 TDS 200 Series Digital Oscilloscope User Manual...
Страница 126: ...Glossary 114 TDS 200 Series Digital Oscilloscope User Manual...
Страница 132: ...Index 120 TDS 200 Series Digital Oscilloscope User Manual...