Glossary
amp-second product
The unit of measure de
fi
ning the maximum amount of pulsed current that can
be measured before the probe core becomes saturated. The amp-second rating
applies only to measurement values between the maximum continuous and
maximum pulse current ratings of the probe. The amp-second rating is equal
to the peak current multiplied by the pulse width at the 50% point.
auto-balance
A process that removes unwanted DC offsets from the TCPA300 and
TCPA400 circuitry to ensure maximum measurement accuracy. This process
is performed during the probe degauss routine.
bucking current
Current fed back by the TCPA300 and TCPA400 to the current probe during
DC and low-frequency AC measurements. Bucking current nulls most of the
magnetic
fi
eld in the probe core, allowing linear DC and AC measurements
simultaneously.
conventional current
fl
ow
The
fl
ow of the positive charge in a conductor. Conventional current
fl
ow is
from positive to negative. The arrows on current probes point in the direction
of conventional current
fl
ow. Conventional current
fl
ow is in the opposite
direction of
electron current
fl
ow
.
degauss
A process of eliminating residual magnetism from the probe core by feeding
a decaying sine wave into the core. The degaussing process should be
performed before clamping the probe around a conductor, whenever the probe
becomes saturated by current overloads, or whenever the probe is exposed
to a magnetic
fi
eld. The term is derived from one of the units of magnetic
fl
ux density.
electron current
fl
ow
The
fl
ow of the electrons in a conductor. Electron current
fl
ow is from
negative to positive. The arrows on current probes point in the opposite
direction of electron current
fl
ow. Electron current
fl
ow is in the opposite
direction of
conventional current
fl
ow
.
fl
ux
The density of a magnetic
fi
eld. As the number of
fl
ux lines increases,
the magnetic
fi
eld strength increases. Current in the conductor under test
generates
fl
ux lines in the probe core by inducing a voltage in the core.
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual
61
Содержание TCP300A
Страница 2: ......
Страница 6: ......
Страница 18: ...Preface xii TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Страница 34: ...Getting Started 16 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Страница 66: ...Troubleshooting and Error Codes 48 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Страница 78: ...Specifications 60 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...