Getting Started
AC Measurements
To measure AC current only, and remove the DC component of the current being
measured, follow the instructions below. These are identical to the instructions for
DC current measurements except that the ampli
fi
er coupling in step 2 is set to AC.
1.
Verify that the oscilloscope input coupling is set to DC, and the input
impedance is set to 50
Ω
. (The NOT TERMINATED INTO 50
Ω
LED on the
ampli
fi
er is lit if impedance is not 50
Ω
.)
2.
Verify that the ampli
fi
er input coupling is AC.
3.
Adjust the ground reference of the oscilloscope to move the trace to the
desired graticule line.
4.
Lock the probe closed without a conductor passing through it, and then press
the ampli
fi
er PROBE DEGAUSS AUTOBALANCE button.
WARNING.
The current probes can be used to measure current on uninsulated
wires. However, the circuit must be de-energized when connecting or removing
the current probe.
5.
Open the probe slide, place the probe around the conductor under test, and
then lock the slide. For correct measurement polarity, make sure the probe
arrow is pointing in the direction of conventional (positive to negative)
current
fl
ow. Reversing the
fl
ow will invert the displayed current waveform
on the oscilloscope.
NOTE.
Even when making AC current measurements, leave the oscilloscope
coupling on DC. Change only the ampli
fi
er coupling to AC. Using the oscilloscope
AC coupling may cause the ampli
fi
er to exceed its output dynamic range.
6.
Adjust the oscilloscope time base and trigger as needed.
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual
15
Содержание TCP300A
Страница 2: ......
Страница 6: ......
Страница 18: ...Preface xii TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Страница 34: ...Getting Started 16 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Страница 66: ...Troubleshooting and Error Codes 48 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Страница 78: ...Specifications 60 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...