Control Summary
Under normal operation, the AC and DC COUPLING LEDs indicate the coupling
mode of the ampli
fi
er. If they alternately
fl
ash after a degauss operation, this
indicates the ampli
fi
er is displaying an error code with the four status LEDs on
the lower-left front panel. (See page 45,
Displaying Error Codes with the Probe
PROBE INPUT Connector
All current probes compatible with the TCPA300 and TCPA400 attach at the
PROBE INPUT connector, which is a multi-pin female connector. Information
about connecting a probe is available. (See page 9,
OUTPUT Connector
The ampli
fi
er current measurement output is accessed at the OUTPUT connector,
which should be connected to the oscilloscope input. Attach one end of a 50
Ω
BNC cable to this connector and the other end to a 50
Ω
vertical input of your
oscilloscope. The output impedance of the ampli
fi
er is 50
Ω
.
To get a direct readout of current on your Tektronix oscilloscope, use the
TEKPROBE interface cable to connect the ampli
fi
er to your oscilloscope.
NOTE.
To obtain accurate measurements, the input impedance of your
oscilloscope must be 50
Ω
. Make sure your ampli
fi
er OUTPUT is connected to an
oscilloscope input using a 50
Ω
BNC cable, and that the oscilloscope input is set
to 50
Ω
impedance. The proper cabling is shown. (See Figure 1 on page 1.)
If your oscilloscope does not have 50
Ω
impedance settings for inputs, you can
place a 50
Ω
feedthrough termination on the oscilloscope input and connect
the ampli
fi
er output cable to the termination. Do not place the feedthrough
termination at the ampli
fi
er end of the BNC connecting cable.
Probe DC Gain Adjust
(located on probes)
After the PROBE DEGAUSS AUTOBALANCE routine has been run, the probe
and ampli
fi
er system will meet all published speci
fi
cations. However, if you want
to improve the tolerance of the system accuracy, or to intentionally offset the
accuracy to make up for total system errors, you can manually adjust the gain of
the probe. See the service manual for adjustment procedures and locations.
NOTE.
You should be careful to note the existing position of the DC Gain
Adjustment before you alter it, so that you may return it to the initial, calibrated
position. By altering the DC Gain Adjustment, you may cause the probe to not
meet the warranted DC Accuracy speci
fi
cation.
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual
21
Содержание TCP300A
Страница 2: ......
Страница 6: ......
Страница 18: ...Preface xii TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Страница 34: ...Getting Started 16 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Страница 66: ...Troubleshooting and Error Codes 48 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...
Страница 78: ...Specifications 60 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...