Repair
Reassembly
1.
Before reassembling the probe body, be sure to verify the following:
a.
The gap between the stationary and moveable core pieces is clean. If necessary, use isopropyl alcohol or a similar
cleaning agent to clean the pieces.
b.
The contacts of the slide switch are clean. (Clean them if necessary.)
c.
The plastic slide assembly is lubricated. If necessary, apply silicone-based grease sparingly to the parts.
2.
Reassemble the probe body by following steps 1 and 2 of the probe head disassembly procedure in reverse. (See page 71,
CAUTION.
To avoid damaging the wires inside the probe, carefully dress the wires in the lower body half to avoid contact with
the slide assembly.
To avoid damaging the switch contacts, exercise care when
fi
tting the slide back into the probe body. Aligning the switch
contacts requires precision.
If the slider does not work smoothly after tightening the two screws, loosen the screws slightly.
TCP0150 Current Probe Instruction Manual
79
Содержание TCP0150
Страница 1: ...TCP0150 20 MHz 150 A AC DC Current Probe Instruction Manual www tektronix com 071 2251 00...
Страница 4: ......
Страница 8: ...Table of Contents iv TCP0150 Current Probe Instruction Manual...
Страница 64: ...Specifications Figure 3 Maximum peak pulse versus pulse width 48 TCP0150 Current Probe Instruction Manual...
Страница 66: ...Specifications Figure 4 Probe dimensions 50 TCP0150 Current Probe Instruction Manual...
Страница 98: ...Repair Figure 5 Replaceable Parts 82 TCP0150 Current Probe Instruction Manual...
Страница 100: ...Repair 84 TCP0150 Current Probe Instruction Manual...