Performance Veri
fi
cation
6.
Set the probe range to 25 A.
7.
Degauss the probe.
8.
Clamp the current probe around the HF
current loop. Verify that the arrow-shaped
indicator on the probe points away from the
pulse generator.
9.
Record the rise time measurement in the
Test Record.
NOTE.
If you measure higher rise times than
expected, check your oscilloscope
fi
rmware
version. (See page 55, Oscilloscope Firmware.)
Early versions of oscilloscope
fi
rmware may
engage the bandwidth limit
fi
lter, which will
yield inaccurate rise time measurements.
10.
Calculate the probe bandwidth using the
measured rise time in the formula shown.
11.
Record the calculated bandwidth value in
the test record.
TCP0150 Current Probe Instruction Manual
65