Performance Veri
fi
cation
Rise Time and Bandwidth
This procedure veri
fi
es that the probe meets the rise time speci
fi
cation. The bandwidth of the probe is then calculated using the
measured probe rise time.
1.
Connect the BNC cable to the output of the
pulse generator.
2.
Connect the other end of the BNC cable to
the HF current loop.
3.
Set the pulse generator output and pulse
width to maximum.
4.
Set the oscilloscope as follows:
Vert: 250 mA/div, Horiz: 10 ns/div
Trigger at 50%, Averaging on (32)
Coupling to DC
Automeasurement to Rise Time
5.
Connect the probe directly to one of the
oscilloscope channels (1–4).
64
TCP0150 Current Probe Instruction Manual