![Tektronix TCP0150 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/tektronix/tcp0150/tcp0150_instruction-manual_1079114020.webp)
Installation
Degaussing the Probe
1.
After the probe is identi
fi
ed by the host
instrument:
The screen display prompts you to
degauss the probe
The multicolor Degauss/AutoZero
status LED on the probe
fl
ashes red
NOTE.
The DC gain and offset are not
guaranteed when this LED
fl
ashes red.
2.
With no conductor in the jaw, lock the
probe slide.
3.
Press the
Degauss/AutoZero
button on
the probe or in the Degauss window on the
host instrument.
4.
The multicolor Degauss/AutoZero status
LED glows green to indicate a successful
degauss routine was run, and that the
probe is in normal operating mode.
4
TCP0150 Current Probe Instruction Manual
Содержание TCP0150
Страница 1: ...TCP0150 20 MHz 150 A AC DC Current Probe Instruction Manual www tektronix com 071 2251 00...
Страница 4: ......
Страница 8: ...Table of Contents iv TCP0150 Current Probe Instruction Manual...
Страница 64: ...Specifications Figure 3 Maximum peak pulse versus pulse width 48 TCP0150 Current Probe Instruction Manual...
Страница 66: ...Specifications Figure 4 Probe dimensions 50 TCP0150 Current Probe Instruction Manual...
Страница 98: ...Repair Figure 5 Replaceable Parts 82 TCP0150 Current Probe Instruction Manual...
Страница 100: ...Repair 84 TCP0150 Current Probe Instruction Manual...