Speci
fi
cations
Speci
fi
cations
The speci
fi
cations in the tables in this section are valid under the following conditions:
The probe has been calibrated at an ambient temperature of 23 °C ±5 °C.
The probe is connected to a host instrument with an input impedance of 1 M
Ω
.
The probe must have a warm-up period of at least 20 minutes and be in an environment that does not exceed the limits
described. (See Table 1).
Speci
fi
cations for the TCP0030A current probe fall into three categories: warranted, typical, and nominal characteristics.
Warranted Characteristics
Warranted characteristics describe guaranteed performance within tolerance limits or certain type-tested requirements. Warranted
characteristics that have checks in the
Performance Veri
fi
cation
section are marked with the
symbol.
Table 1: Warranted electrical characteristics
Characteristic
Description
DC gain accuracy
<3% (typical <1% at +23 °C, ±5 °C)
Rise time (10% to 90%)
≤
2.92 ns
Bandwidth
DC to 120 MHz
TCP0030A Current Probe Instruction Manual
45
Содержание TCP0030A
Страница 1: ...TCP0030A 120 MHz 30 Amp AC DC Current Probe ZZZ Instruction Manual xx www tek com 071 3006 01...
Страница 4: ......
Страница 8: ...Table of Contents iv TCP0030A Current Probe Instruction Manual...
Страница 20: ...Preface xvi TCP0030A Current Probe Instruction Manual...
Страница 68: ...Specifications Figure 2 Typical input impedance versus frequency 48 TCP0030A Current Probe Instruction Manual...
Страница 69: ...Specifications Figure 3 Maximum peak pulse versus pulse width TCP0030A Current Probe Instruction Manual 49...
Страница 71: ...Specifications TCP0030A Current Probe Instruction Manual 51...