Installation
Degauss the Probe
1.
After the probe is identi
fi
ed by the host
instrument, the screen display prompts you
to run a degauss routine.
The multicolor Degauss/AutoZero status
LED on the probe also indicates that a
degauss is needed, by
fl
ashing red.
2.
To degauss the probe,
fi
rst verify that the
probe jaw is closed, and then press the
Degauss/AutoZero
button on the probe
or in the Degauss window on the host
instrument.
NOTE.
The DC gain and offset errors are not
guaranteed when this LED
fl
ashes red.
3.
The multicolor Degauss/AutoZero status
LED glows green to indicate a successful
degauss routine was run, and that the
probe is in normal operating mode.
TCP0030A Current Probe Instruction Manual
3