Performance Veri
fi
cation
Rise Time and Bandwidth
This procedure veri
fi
es that the probe meets the rise time speci
fi
cation on both of the current ranges. The bandwidth of the
probe is then calculated using the measured probe rise time.
1.
Connect the probe to any channel (1–4).
2.
Connect the BNC cable to the output of
the pulse generator, using a BNC-to-SMA
adapter if necessary.
3.
Connect the other end of the BNC cable to
the HF current loop.
4.
Set the pulse generator output and pulse
width to maximum (0 dB, 100 ns duration).
5.
Set the oscilloscope:
Vertical sensitivity to 200 mA/div
Horizontal to 2 ns/div
Trigger at 50%
Averaging on (32)
Coupling to DC
To measure Rise Time
TCP0030A Current Probe Instruction Manual
61