Performance Tests
Performance Verification
4Ć14
TableĂ4Ć1:ăRequired Test Equipment
Item Numberand
Description
Minimum Requirements
Example
Purpose
1 Attenuator, 5X
1
Ratio: 5X; Impedance 50
W
;
Connectors: female BNC input,
male BNC output
Tektronix part number
011-0060-02
Signal attenuation
2 Attenuator, 10X
Ratio: 10X; Impedance 50
W
;
Connectors: female BNC input,
male BNC output
Tektronix part number
011-0059-02
Signal attenuation
3 Termination,
50
W
(two reĆ
quired)
Impedance 50
W
; Connectors:
female BNC input, male BNC
output
Tektronix part number
011-0049-01
Signal
interconnection
4 Termination,
75
W
Impedance 75
W
; Connectors:
female BNC input, male BNC
output
Tektronix part number
011-0102-01
Signal
interconnection,
video
5 Cable, Precision
Coaxial
50
W
, 36 in, maleĆtoĆmale BNC
connectors
Tektronix part number
012-0482-00
Signal
interconnection
6 Cable, Coaxial
75
W
, 36 in, maleĆtoĆmale BNC
connectors
Tektronix part number
012-1338-00
Signal
interconnection,
video
7 Coupler, DualĆ
Input
FemaleĆBNCĆtoĆdualĆmaleĆBNC
Tektronix part number
067Ć0525Ć02
Signal
interconnection
8 Generator, LevĆ
eled Sine Wave
200 kHz to 250 MHz; Variable
amplitude from 5ĂmV to 4 V
pĆp
into 50
W
Wavetek 9100 Universal
Calibration System with
Oscilloscope Calibration
Module (Option 250)
Trigger and
bandwidth checks
9 Generator, Time
Mark
Variable marker frequency from
10 ms to 10 ns; accuracy within
25 ppm
Module (Option 250)
Timing accuracy
checks
10 Generator, Pulse High Amplitude pulse with variable
amplitude of 60 V to 100 V
Gain accuracy
checks
11 Generator, DC
Calibration
DC voltage levels from 100 mV to
10 V
Gain accuracy
checks
12 Probe, 10X, inĆ
cluded with this
instrument
Standard accessory probe
TEKTRONIX P6109B
Signal
interconnection
13 Digital
Multimeter
DC voltage to 10ĂV
TEKTRONIX DMĂ2510
Power Supply
Adjustment
14 Generator, Video
Signal
Provides NTSC compatible
outputs
TEKTRONIX TSG 100
Video trigger check
15 Adjustment Tool
1
0.1 inch hex on both ends
GC Electronics #8606
Internal adjustments
1
This item is used only when performing the Adjustment Procedures.
Содержание TAS 455
Страница 4: ......
Страница 10: ...vi...
Страница 13: ...TAS 455 and TAS 465 Service Manual ix...
Страница 14: ...x...
Страница 16: ...xii...
Страница 17: ...TAS 455 and TAS 465 Service Manual xiii Safety Summary H H H H Symbols and Terms...
Страница 20: ...xvi...
Страница 21: ...TAS 455 and TAS 465 Service Manual xvii Preface...
Страница 22: ...xviii H Reference H Instruction H XYZs of Analog and Digital Oscilloscopes...
Страница 23: ......
Страница 24: ...TAS 455 and TAS 465 Service Manual 1 1 Product Description H H 0 H H H H H H H H...
Страница 27: ...1 4...
Страница 39: ...1 16...
Страница 44: ......
Страница 53: ......
Страница 54: ...TAS 455 and TAS 465 Service Manual 3 1 Circuit Description Logic Conventions Module Interconnection...
Страница 62: ......
Страница 74: ...Brief Performance Checks Performance Verification 4 12...
Страница 92: ...4 30...
Страница 93: ......
Страница 111: ...5 18...
Страница 115: ...Adjustment Procedures Adjustment Procedures 5 22 J205 CAL DIS Figure 5 17 Location of J205 CAL DIS on the CPU Board...
Страница 116: ......
Страница 120: ...6 4...
Страница 124: ...6 8...
Страница 141: ...Removal and Installation Procedures TAS 455 and TAS 465 Service Manual 6 25 H H H 0 0 0 R 0 0 R...
Страница 149: ...Removal and Installation Procedures TAS 455 and TAS 465 Service Manual 6 33 CRT Installation 1 0 0 1 2 1 0 2 0 0...
Страница 156: ...Removal and Installation Procedures Maintenance 6 40...
Страница 167: ...TAS 455 and TAS 465 Service Manual 6 51 Repackaging Information...
Страница 168: ...6 52...
Страница 169: ......
Страница 172: ......
Страница 178: ...8 6...
Страница 204: ...8 32...
Страница 205: ......
Страница 208: ......
Страница 209: ...0 0 0 0 0 0 0 0 0 0 0 0 1 0 0 1 1...
Страница 210: ...10 2...
Страница 213: ...10 5...
Страница 215: ...10 7...
Страница 216: ...10 8...
Страница 220: ...10 12...
Страница 222: ...10 14...
Страница 224: ...10 16...
Страница 225: ......
Страница 226: ......