TAS 200 Series Adjustment Procedures
26
Bench Test Instruments and Handheld Oscilloscopes Basic Service
Following the warm-up period, preset the oscilloscope to the settings listed
below.
Table 20: Oscilloscope initial settings
Control
Setting
INTENSITY
Visible display
CURSOR ON/OFF
Off
VERTICAL
CH 1 (CH 2) POSITION
Midrange
MODE
CH1 (CH2)
CH 1 (CH 2) VOLTS/DIV
10 mV
CH 1 (CH 2) VOLTS/DIV VARIABLE
CAL
CH 1 (CH 2) GND
In
HORIZONTAL
POSITION
Midrange
X10 MAG
Out
X-Y
Out
SWP UNCAL
Out
TRIGGER
LEVEL
Midrange
MODE
AUTO
COUPLING
AC
SOURCE
CH 1 (CH 2)
SLOPE
Rising (push
switch out)
Initial Settings