
TAS 200 Series Performance Verification
Bench Test Instruments and Handheld Oscilloscopes Basic Service
15
Horizontal Check
The following checks verify the horizontal accuracy of your oscilloscope.
To check time base accuracy, perform the following steps.
1.
Use the 50
Ω
precision coaxial cable to connect the output of the time mark
generator to the 50
Ω
termination; then connect the 50
Ω
termination to the
TAS 200 series instrument CH 1 input. See Figure 4 below.
Time Mark
Generator
50
Ω
Termination
Precision cable
Figure 4: Timing check setup
2.
Set up the oscilloscope as follows:
VERTICAL MODE
CH1
CH 1 AC-DC
DC
CH 1 VOLTS/DIV
0.5 V
CH 1 GND
Out (release)
HORIZONTAL SEC/DIV
.1
s
TRIGGER MODE
AUTO
TRIGGER COUPLING
DC
3.
Set up the time mark generator to produce 0.1
s markers.
4.
Center the time marks vertically on the oscilloscope display.
5.
Position the rising edge of the second time mark to the second vertical
graticule line of the oscilloscope display.
6.
Verify that the time mark to graticule accuracy over the center eight divisions
is within the limits shown for each HORIZONTAL SEC/DIV setting listed
in Table 15.
Time Base Accuracy
Содержание TAS 200
Страница 1: ...Instructions Manual TAS 200 Series Oscilloscopes 070 9855 01 www tektronix com ...
Страница 2: ......
Страница 4: ...Table of Contents ii Bench Test Instruments and Handheld Oscilloscopes Basic Service ...
Страница 6: ...TAS 200 Series Oscilloscopes 2 Bench Test Instruments and Handheld Oscilloscopes Basic Service ...