TAS 200 Series Performance Verification
Bench Test Instruments and Handheld Oscilloscopes Basic Service
13
NOTE
. At lower VOLTS/DIV settings, you might need to set TRIGGER COU-
PLING to HF REJ to minimize noise or double triggering.
Table 14: DC coupled bandwidth
Volts/Div
50 kHz
reference amplitude
Bandwidth
minimum amplitude
TAS 220
bandwidth
TAS 250
bandwidth
1 mV
6 divisions
4.2 divisions
10 MHz
15 MHz
2 mV
6 divisions
4.2 divisions
10 MHz
15 MHz
5 mV
6 divisions
4.2 divisions
20 MHz
50 MHz
10 mV
6 divisions
4.2 divisions
20 MHz
50 MHz
20 mV
6 divisions
4.2 divisions
20 MHz
50 MHz
50 mV
6 divisions
4.2 divisions
20 MHz
50 MHz
100 mV
6 divisions
4.2 divisions
20 MHz
50 MHz
200 mV
6 divisions
4.2 divisions
20 MHz
50 MHz
500 mV
6 divisions
4.2 divisions
20 MHz
50 MHz
1 V
4 divisions
2.8 divisions
20 MHz
50 MHz
4.
Repeat steps 1 through 3 for CH 2.
To check DC balance accuracy, perform the following steps.
1.
Set up the oscilloscope as follows:
VERTICAL MODE
CH1 (CH2)
CH 1 (CH 2) GND
In
CH 1 (CH 2) VARIABLE
CAL
2.
Verify that the oscilloscope trace shift is less than
±
0.5 divisions when
rotating the CH 1 (CH 2) VOLTS/DIV control between the 5 mV and 5 V
settings.
3.
Verify that the oscilloscope trace shift is less than two divisions when the
CH 1 (CH 2) VOLTS/DIV control is changed from the 1 mV to the 2 mV
setting.
4.
Repeat steps 1 through 3 above for CH 2.
DC Balance