TAS 200 Series Performance Verification
Bench Test Instruments and Handheld Oscilloscopes Basic Service
11
Table 13: DC gain and displayed signal accuracy
TAS 200 volts/div scale
Generator output
Displayed signal accuracy
1 mV
5 mV
4.75 to 5.25 div
2 mV
10 mV
4.75 to 5.25 div
5 mV
20 mV
3.88 to 4.12 div
10 mV
50 mV
4.85 to 5.15 div
20 mV
0.1 V
4.85 to 5.15 div
50 mV
0.2 V
3.88 to 4.12 div
100 mV
0.5 V
4.85 to 5.15 div
200 mV
1 V
4.85 to 5.15 div
500 mV
2 V
3.88 to 4.12 div
1 V
5 V
4.85 to 5.15 div
2 V
10 V
4.85 to 5.15 div
5 V
20 V
3.88 to 4.12 div
4.
Set the calibration generator output to 5 mV.
5.
Return the oscilloscope CH 1 (CH 2) VOLTS/DIV control to 1 mV.
6.
Rotate the oscilloscope CH 1 (CH 2) VERTICAL VARIABLE control
counterclockwise off of the CAL position until the amplitude of the
displayed waveform is reduced to two divisions or less.
7.
Set the oscilloscope CH 1 (CH 2) VERTICAL VARIABLE control to the
CAL position.
8.
Disconnect the test setup from the oscilloscope.
9.
Repeat steps 1 through 8 for CH 2.