Performance Verification Procedures
SDA 601 Serial Digital Analyzer
3- 17
e.
Set the following parameters in the SDA601 Utility:RS232 Setup
submenu:
H
Baud Rate = 9600
H
Set Parity = None
f.
Press
Shift
, then
Analyze
on the SDA601.
Verify
that the PC receives a
Analysis report similar to the one shown in Figure 2--5.
18.
Return Loss
Specification:
≥
15 dB from 1 MHz to 270 MHz, power On.
a.
Connect a precision 50
Ω
cable from the spectrum analyzer TG output to
the RF input of the RF bridge.
b.
Connect a precision 50
Ω
cable from the spectrum analyzer RF input to
the RF output of the RF bridge.
c.
Connect a
75
Ω
male-to-male BNC adapter to the RF bridge “device
under test” connector. The equipment setup should now resemble the
arrangement illustrated in Figure 3--14.
Figure 3- 14: Initial Return Loss Setup
Содержание SDA601
Страница 4: ......
Страница 10: ......
Страница 15: ...Specifications...
Страница 16: ......
Страница 21: ...Operating Information...
Страница 22: ......
Страница 29: ...Operating Information SDA 601 Serial Digital Analyzer 2 7 Figure 2 5 A Printed Analysis Report...
Страница 49: ...Performance Verification...
Страница 50: ......
Страница 69: ...Adjustment Procedures...
Страница 70: ......
Страница 78: ...Adjustment Procedures 4 8 SDA 601 Serial Digital Analyzer...
Страница 79: ...Maintenance...
Страница 80: ......
Страница 93: ...Replaceable Electrical Parts...
Страница 94: ......
Страница 104: ......
Страница 105: ...Diagrams...
Страница 106: ......
Страница 108: ...Diagrams and Circuit Board Illustrations 7 2 SDA 601 Serial Digital Analyzer Data Path Diagram...
Страница 111: ......
Страница 122: ......
Страница 123: ...Replaceable Mechanical Parts...
Страница 124: ......
Страница 129: ...Index...
Страница 130: ......